Product Description:
Gradient PCR Thermal Cycler (Laboratory, Color Touch Screen)
This is a laboratory gradient PCR thermal cycler. It rapidly heats and cools the sample block to run the denaturation, annealing and extension cycles of DNA amplification automatically. The gradient function tests several annealing temperatures in a single run, making it faster to optimise PCR conditions for a new primer pair. A color touch screen handles programme setup and run monitoring — suited to molecular biology labs, CDC facilities, universities, clinical laboratories and quality-inspection institutes.
 
 
Product Highlights:
· Factory Direct Supply: supplied directly by HIPROVE, a laboratory-equipment manufacturer, with stable availability for importers, distributors and lab-equipment channels placing bulk and repeat orders;
· Gradient Function Saves Experiment Time: several annealing temperatures can be compared in one run instead of repeated trial runs, cutting reagent, consumable and labour costs — a practical configuration for method development and primer optimisation;
· Color Touch Screen Programming: an intuitive graphical interface for setting temperatures, hold times, cycle counts and programme steps, with clear run status and a short learning curve for new users;
· Flip-top Heated Lid on a Compact Benchtop Body: the lift-up lid and multi-well sample block sit in a small footprint, easy to place on a routine lab bench and to run alongside other instruments;
· Broad Application: covers routine PCR amplification and condition optimisation in research, teaching, disease control, inspection and quarantine, food and environmental testing — a versatile unit for distribution channels;
· OEM/ODM Support: private labelling, colour and configuration options available for regional distributors building their own brand;
· Flexible Order Quantity: a minimum order quantity of 1 piece supports samples, demonstration units and small trial orders with a low entry barrier for first-time buyers;
· Export Ready: standard export packaging for long-distance shipment, with stable lead times and after-sales coordination for overseas buyers.
 
 
Working Principle:
A solid-state Peltier thermoelectric module drives heating and cooling of the sample block by reversing the current direction; temperature sensors and a closed-loop controller hold the block at the set point. The run programme cycles through denaturation, annealing and extension, with temperatures, hold times and cycle counts set on the touch screen. In gradient mode a temperature spread is established across the block so that different well positions correspond to different annealing temperatures, allowing the best condition to be identified in a single experiment. The heated lid keeps tube caps hot so that evaporated liquid cannot condense on a cooler cap and change the reaction volume — removing the need for a mineral-oil overlay.
 
 
Operation Summary:
1. Place the unit on a stable bench free of strong air currents and vibration, connect power and allow the warm-up recommended in the manual;
2. Load PCR tubes or plates containing the reaction mix into the sample block, checking that the consumables match the block format and are fully seated;
3. Create or recall a programme on the touch screen, setting temperatures, hold times, cycle counts and lid status; define a gradient temperature range when optimising conditions;
4. Close and secure the heated lid, start the run, and follow progress and status on the screen;
5. After the run, remove samples for downstream analysis, clean the block surface, power down and record the run.
 
 
Packaging & Shipping:
Standard export packaging suitable for long-distance shipment; carton dimensions, gross/net weight and delivery time are confirmed per order.
 
 
Important Notes:
1. Confirm the supply voltage matches the nameplate and complete the warm-up and any required calibration before use;
2. Use only PCR consumables matched to the sample block, and ensure caps are properly sealed to avoid evaporation or condensation;
3. Avoid opening the lid during a run; the hot block and heated lid present a scald risk — use protection when handling samples;
4. Keep the block clean and dry, prevent liquid or dust entering the housing, and service the unit as directed in the manual;
5. This instrument is for nucleic-acid amplification; results must be interpreted by trained personnel together with the reagents and workflow, and are not a stand-alone diagnostic result.

 